MINBAR X-ray burst light curves / method

Generating and preparing light curves

All light curves were extracted for the full bandpass and with a resolution of 1 s. The basis for the extraction was a pre-determined list of burst onset times. Standard, a time frame of 50 s before till 300 s after the onset time was extracted. For incidental long bursts (from particularly GX 17+2), these time frames were chosen on a burst-by-burst case.

RXTE

For RXTE, the light curves were extracted from the 'standard-1' data (0.125 s resolution, no energy resolution, PCUs resolved). For incidental cases, data in this mode were lacking and event-mode data were employed instead. The light curves were normalized to the number of active PCUs and the collimator response. The collimator response was calculated as one minus the off-axis angle of the source in degrees. A background level was subtracted that is the average of the flux between -22 and -7 s with respect to the burst onset time. No deadtime correction was applied. Note that these light curves are subtracted for background emission as well as source persistent emission.

WFC

For WFC, the light curves were generated from PSF fitting 1-s images between -50 and +300 s from burst onset. The source position and PSF shape is fixed in these fits, only the flux is left free. The source position was determined in an iterative manner: 1) the burst time and duration is identified in a light curve of the complete detector; 2) an image is reconstructed for this time frame; 3) the burst source is identified in this image; 4) an imaged light curve is generated for this source using the initial position; 5) this light curve is employed to determine the optimum time frame for the best signal-to-noise ratio; 6) a new image is determined for this optimum time frame; 7) the most accurate source position is determined from this image. Note that these light curves are not subtracted for source persistent emission.

JEM-X

The light curves were obtained from Duncan. I am unaware of the manner in which they were generated, but they seem to be generated in two different manners, since they seem to have two different units: one in c/s/collimator-response and the other in c/s/cm2. There are 268 bursts for which both JEM-X1 and JEM-X2 were active. The signals from both units were combined and divided by 2 to keep the same normalization.

Fitting light curves

First, the light curves needed to be normalized to the same flux unit. For the PCA data, the light curves were re-normalized from c/s/PCU to c/s/cm2 by division through 1300. For JEM-X, some light curves needed to be re-normalized from c/s to c/s/cm2 by division by 100. Times were redefined to be measured with respect to input burst onset time.

The following tasks were carried out

Calibration

The data need to be calibrated between the three instruments. This can be done in two ways:

Exceptions

The following bursts have been excluded from the analysis at a late stage:

RXTE-PCA

BeppoSAX-WFC


Jean in 't Zand (SRON, 30-Aug-16